They are instructed in the scanner firmware help which is accessible through service terminal. Wait that the unit starts. Check that there is no mechanical reason which prevents linear movement for example image plate or protective covers. Turn the unit on and start S2terminal This has been found as a root cause to many image quality related complaints with the new Optime. digora optime drivers

Uploader: Mojinn
Date Added: 19 August 2011
File Size: 30.98 Mb
Operating Systems: Windows NT/2000/XP/2003/2003/7/8/10 MacOS 10/X
Downloads: 97145
Price: Free* [*Free Regsitration Required]





After calibration quit from service mode and turn the unit off and again on. Double-click on Digora from www. If after reassembly safety switch error occurs the light cover might be raised from the chassis.

DIGORA Optime DXR-60 Field Service Manual

The -a flag tells Digora to change focus to itself. However it is not always digoar display the error code. Checking linear movement errors It is recommended first to check that the motor cable is connected to the motor and K board and that it is not broken motor is not rotating.

It's a community-based project which helps to repair anything. High voltage can be set on when the light beam of opto safety switch is blocked. Instead of the tool you can also use torque wrench. Turn the unit on and start S2terminal The opime should now start without errors.

Unit does not react to any other inputs than those given by the user through service terminal. Detach the connectors and remove the covers. Insert the main cover Open the covers and check that the cable is properly connected to K board and user interface board. Check the motor connector.

digora optime drivers

The laser can be activated if the light beam in the optic safety switch is blocked and the laser is switched on from the service terminal. Replacement of the plates needed. Attach laser cable as in the image below. Attach the back cover If the cable reconnecting does not help, then ooptime user interface panel.

Laser is faulty or laser connector is not connected.

digora optime drivers

K is not a spare part. Corrected from serial SL Service: The power adapter has accessible conductive part DC plug but this is newer opened or serviced and it is always measured when delivered from factory. Blue led at K board is used as the light source.

Open Dental Software - Digora Soredex Optime Bridge

Digora for Windows 2. Service terminal commands and functions are not explained in this manual. Replaced K board, calibrated and endurance test performed.

digora optime drivers

After testing, repair or adjustment, check that the unit is restored to the conditions necessary for normal use. Check that it goes to connector and that the slot on the pcb goes to the holder marked A.

Release spring B by pressing it. The chassis is locked to the hole in the clip of the light cover. Using latest FW and core versions provides the unit with all features and improvements Firmware and core upgrading is done using s2terminal —program.

Check that opto actuators in the carriage go through the opto slots without touching the optos. Now it should start without errors.